User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: B&K 4011: 5 MHz Function Generator DATE: 06-Aug-97 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 42 NUMBER OF LINES: 476 CONFIGURATION: HP 5335A CONFIGURATION: HP 34401A STANDARD: Fluke A55 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 STD Fluke A55 1.004 IEEE [@5335][TERM LF] 1.005 IEEE [@535]IN; 1.006 HEAD INITIAL CONDITIONS 1.007 DISP Set the UUT as follows: 1.007 DISP RANGE 5 1.007 DISP function sine 1.007 DISP OUTPUT LEVEL full ccw 1.007 DISP DUTY CYCLE OFF 1.007 DISP CMOS LEVEL OFF 1.008 DISP Set the UUT as follows: 1.008 DISP DC OFFSET OFF 1.008 DISP -20 db OFF 1.008 DISP FREQUNCY COARSE/FINE full ccw 1.009 HEAD {} 1.010 HEAD {* Frequency Accuracy *} 1.011 JMP 2.001 1.012 EVAL 2.001 DISP Connect the UUT OUTPUT BNC via a 50 ohm feedthrough 2.001 DISP termination to INPUT A on the 5335A counter. 2.002 DISP Set the UUT output level control midway. 2.003 DISP Adjust the UUT FREQUENCY controls for a UUT readout 2.003 DISP of 000.5 Hz. 2.004 IEEE [@5335]FU1,AX1,AU1; 2.005 IEEE [@5335][I] 2.006 IEEE [@5335][T0][I] 2.007 ACC .500H 0.0010% 2.008 MEME 2.009 MEMC H 0.1000U #! Test Tol 0.1, Sys Tol 5e-006, TUR 20000.000 (>= 4.00). 3.001 DISP Adjust the UUT FREQUENCY controls for a UUT readout 3.001 DISP of 005.0 Hz. 3.002 IEEE [@5335][T0][I] 3.003 ACC 5.000H 0.0010% 3.004 MEME 3.005 MEMC H 0.10005U #! Test Tol 0.10005, Sys Tol 5e-005, TUR 2001.000 (>= 4.00). 4.001 DISP Set the UUT frequency RANGE to 50. 4.002 DISP Adjust the UUT FREQUENCY controls for a UUT readout 4.002 DISP of 0005 Hz. 4.003 IEEE [@5335][T0][I] 4.004 ACC 5.000H 0.0010% 4.005 MEME 4.006 MEMC H 1.00005U #! Test Tol 0.10005, Sys Tol 5e-005, TUR 2001.000 (>= 4.00). 5.001 DISP Adjust the UUT FREQUENCY controls for a UUT readout 5.001 DISP of 0050 Hz. 5.002 IEEE [@5335][T0][I] 5.003 ACC 50.000H 0.0010% 5.004 MEME 5.005 MEMC H 1.0005U #! Test Tol 1.0005, Sys Tol 0.0005, TUR 2001.000 (>= 4.00). 6.001 DISP Set the UUT frequency RANGE to 500. 6.002 DISP Adjust the UUT FREQUENCY controls for a UUT readout 6.002 DISP of 0050 Hz. 6.003 IEEE [@5335][T0][I] 6.004 ACC 50.000H 0.0010% 6.005 MEME 6.006 MEMC H 1.0005U #! Test Tol 1.0005, Sys Tol 0.0005, TUR 2001.000 (>= 4.00). 7.001 DISP Adjust the UUT FREQUENCY controls for a UUT readout 7.001 DISP of 0500 Hz. 7.002 IEEE [@5335][T0][I] 7.003 ACC 500.00H 0.0010% 7.004 MEME 7.005 MEMC H 1.005U #! Test Tol 1.005, Sys Tol 0.005, TUR 201.000 (>= 4.00). 8.001 DISP Set the UUT frequency RANGE to 5K. 8.002 DISP Adjust the UUT FREQUENCY controls for a UUT readout 8.002 DISP of 0500 Hz. 8.003 IEEE [@5335][T0][I] 8.004 ACC 500.00H 0.0010% 8.005 MEME 8.006 MEMC H 1.005U #! Test Tol 1.005, Sys Tol 0.005, TUR 201.000 (>= 4.00). 9.001 DISP Adjust the UUT FREQUENCY controls for a UUT readout 9.001 DISP of 5000 Hz. 9.002 IEEE [@5335][T0][I] 9.003 ACC 5000.00H 0.0010% 9.004 MEME 9.005 MEMC H 1.05U #! Test Tol 1.05, Sys Tol 0.05, TUR 21.000 (>= 4.00). 10.001 DISP Set the UUT frequency RANGE to 50K. 10.002 DISP Adjust the UUT FREQUENCY controls for a UUT readout 10.002 DISP of 05.00 kHz. 10.003 IEEE [@5335][T0][I] 10.004 MEM/ 1000 10.005 ACC 5.000kH 0.0010% 10.006 MEME 10.007 MEMC kH 0.01005U #! Test Tol 0.01005, Sys Tol 5e-005, TUR 201.000 (>= 4.00). 11.001 DISP Adjust the UUT FREQUENCY controls for a UUT readout 11.001 DISP of 50.00 kHz. 11.002 IEEE [@5335][T0][I] 11.003 MEM/ 1000 11.004 ACC 50.000kH 0.0010% 11.005 MEME 11.006 MEMC kH 0.0105U #! Test Tol 0.0105, Sys Tol 0.0005, TUR 21.000 (>= 4.00). 12.001 DISP Set the UUT frequency RANGE to 500K. 12.002 DISP Adjust the UUT FREQUENCY controls for a UUT readout 12.002 DISP of 050.0 kHz. 12.003 IEEE [@5335][T0][I] 12.004 MEM/ 1000 12.005 ACC 50.000kH 0.0010% 12.006 MEME 12.007 MEMC kH 0.1005U #! Test Tol 0.1005, Sys Tol 0.0005, TUR 201.000 (>= 4.00). 13.001 DISP Adjust the UUT FREQUENCY controls for a UUT readout 13.001 DISP of 500.0 kHz. 13.002 IEEE [@5335][T0][I] 13.003 MEM/ 1000 13.004 ACC 500.000kH 0.0010% 13.005 MEME 13.006 MEMC kH 0.105U #! Test Tol 0.105, Sys Tol 0.005, TUR 21.000 (>= 4.00). 14.001 DISP Set the UUT frequency RANGE to 5M. 14.002 DISP Adjust the UUT FREQUENCY controls for a UUT readout 14.002 DISP of 0500 kHz. 14.003 IEEE [@5335][T0][I] 14.004 MEM/ 1000 14.005 ACC 500.000kH 0.0010% 14.006 MEME 14.007 MEMC kH 1.005U #! Test Tol 1.005, Sys Tol 0.005, TUR 201.000 (>= 4.00). 15.001 DISP Adjust the UUT FREQUENCY controls for a UUT readout 15.001 DISP of 5000 kHz. 15.002 IEEE [@5335][T0][I] 15.003 MEM/ 1000 15.004 ACC 5000.00kH 0.0010% 15.005 MEME 15.006 MEMC kH 1.05U #! Test Tol 1.05, Sys Tol 0.05, TUR 21.000 (>= 4.00). 16.001 DISP Disconnect all test equipment. 16.002 HEAD {} 16.003 HEAD {* Output Tests *} 16.004 JMP 17.001 16.005 EVAL 17.001 DISP Connect the UUT OUTPUT to the DVM. DO NOT use a 50 ohm 17.001 DISP termination. 17.002 DISP Set the UUT frequency RANGE to 5K. 17.003 DISP Adjust the UUT FREQUENCY controls for a UUT readout 17.003 DISP of 1000 Hz. 17.004 DISP Set the UUT output level control full CW. 17.005 IEEE [@34401]*RST 17.006 IEEE [@34401] MEAS:VOLT:AC?[I] 17.007 MEM* 2.828 17.008 MEME >1 17.009 MATH m[1]=m[1]/2 17.010 ACC 10 20.000Vp-p 0.06% 0.03/ 17.011 MEME 17.012 MEMC Vp-p 10.00% open #! Test Tol 20, Sys Tol 0.015, TUR 1333.333 (>= 4.00). 18.001 DISP Connect the UUT OUTPUT to the DVM through the 50 ohm 18.001 DISP termination. 18.002 IEEE [@34401] MEAS:VOLT:AC?[I] 18.003 MEM* 2.828 18.004 MEME 18.005 MEME <1 18.006 MEMC Vp-p 10.00% term. #! T.U.R. not calculated because System Uncertainty not available. 19.001 HEAD {} 19.002 HEAD {* -20dB Test *} 19.003 IEEE [@34401]CALC:STAT ON 19.003 IEEE [@34401]CALC:FUNC DB 19.004 IEEE [@34401]READ?[I] 19.005 DISP Set the UUT -20dB attenuator to ON. 19.006 IEEE [@34401]READ?[I] 19.007 ACC -20.00D 0.001U 19.008 MEME 19.009 MEMC D 1.00U #! Test Tol 1, Sys Tol 0.001, TUR 1000.000 (>= 4.00). 20.001 IEEE [@34401]CALC:STAT OFF 20.002 DISP Disconnect all test equipment. 20.003 HEAD {} 20.004 HEAD {* DC Offset Test *} 20.005 IEEE [@34401]*RST 20.006 DISP Connect the UUT OUTPUT to the DVM. Do not use a 20.006 DISP termination. 20.007 DISP Set the UUT as follows: 20.007 DISP OUTPUT LEVEL midway 20.007 DISP -20dB OFF 20.007 DISP DC OFFSET ON 20.008 DISP Rotate the DC OFFSET control fully CCW. 20.009 IEEE [@34401]MEAS:VOLT:DC?[I] 20.010 ACC 10 -10.00V .0035% .0005/ 20.011 MEME 20.012 MEMC V 100.00% #! Test Tol 10, Sys Tol 0.0004, TUR 25000.000 (>= 4.00). 21.001 DISP Rotate the DC OFFSET control fully CW. 21.002 IEEE [@34401]MEAS:VOLT:DC?[I] 21.003 ACC 10 10.00V .0035% .0005/ 21.004 MEME 21.005 MEMC V 100.00% #! Test Tol 10, Sys Tol 0.0004, TUR 25000.000 (>= 4.00). 22.001 DISP Set the UUT as follows: 22.001 DISP OUTPUT LEVEL full ccw 22.001 DISP DC OFFSET OFF 22.002 DISP Disconnect all test equipment. 22.003 HEAD {} 22.004 HEAD {* Sine Wave Tests *} 22.005 HEAD {* Frequency Response *} 22.006 HEAD {Referenced to 1 kHz} 22.007 HEAD Sine Wave Tests 22.008 JMP 23.001 22.009 EVAL 23.001 DISP Set the UUT as follows: 23.001 DISP RANGE 5K 23.001 DISP FREQUENCY 1000 23.002 DISP Connect the UUT OUTPUT BNC to the 34401A input. 23.003 IEEE [@34401]CONF:VOLT:AC[GTL] 23.004 DISP Adjust the UUT OUTPUT LEVEL control for a 3.00V 23.004 DISP on the DMM. 23.005 IEEE [@34401]MEAS:VOLT:AC?[I] 23.006 MEME >20 23.007 DISP Set the UUT 20 dB to ON. 23.008 DISP Connect the Fluke A55 Thermal Converter to the 23.008 DISP UUT OUTPUT BNC. 23.009 DISP Connect the Thermal Converter output to the DVM input. 23.010 DISP Set the UUT 20 dB to OFF. 23.011 HEAD [27][91]7m[27][91]5mPlease Wait, Measurement in Progress 23.012 IEEE [@34401][D3]MEAS:VOLT:DC?[I] 23.013 HEAD * Sine Wave Tests * #23.008 MEM* 1000 23.014 MEME >1 23.015 MATH m[19] = m[1] / m[20] 23.016 DISP Set the UUT as follows: 23.016 DISP 20 dB ON 23.016 DISP RANGE 5 23.017 DISP Set the FREQUENCY controls full CW. 23.018 DISP Set the UUT 20 dB to OFF. 23.019 HEAD [27][91]7m[27][91]5mPlease Wait, Measurement in Progress 23.020 IEEE [@34401][D3]MEAS:VOLT:DC?[I] 23.021 HEAD * Sine Wave Tests * #23.016 MEM* 1000 23.022 MATH mem = mem / m[19] 23.023 MATH mem = mem - m[20] 23.024 MEM/ 2 23.025 MATH mem = mem + m[20] 23.026 MEME 23.027 MEME <20 23.028 MEMC V 5.00% 6H #! T.U.R. not calculated because System Uncertainty not available. 24.001 DISP Set the UUT as follows: 24.001 DISP 20 dB ON 24.001 DISP RANGE 50 24.002 DISP Adjust the UUT FREQUENCY controls for a UUT readout 24.002 DISP of 0050 Hz. 24.003 DISP Set the UUT 20 dB to OFF. 24.004 HEAD [27][91]7m[27][91]5mPlease Wait, Measurement in Progress 24.005 IEEE [@34401][D3]MEAS:VOLT:DC?[I] 24.006 HEAD * Sine Wave Tests * #24.007 MEM* 1000 24.007 MATH mem = mem / m[19] 24.008 MATH mem = mem - m[20] 24.009 MEM/ 2 24.010 MATH mem = mem + m[20] 24.011 MEME 24.012 MEME <20 24.013 MEMC V 5.00% 50H #! T.U.R. not calculated because System Uncertainty not available. 25.001 DISP Set the UUT as follows: 25.001 DISP 20 dB ON 25.001 DISP RANGE 500 25.002 DISP Adjust the UUT FREQUENCY controls for a UUT readout 25.002 DISP of 0500 Hz. 25.003 DISP Set the UUT 20 dB to OFF. 25.004 HEAD [27][91]7m[27][91]5mPlease Wait, Measurement in Progress 25.005 IEEE [@34401][D3]MEAS:VOLT:DC?[I] 25.006 HEAD * Sine Wave Tests * #25.007 MEM* 1000 25.007 MATH mem = mem / m[19] 25.008 MATH mem = mem - m[20] 25.009 MEM/ 2 25.010 MATH mem = mem + m[20] 25.011 MEME 25.012 MEME <20 25.013 MEMC V 5.00% 500H #! T.U.R. not calculated because System Uncertainty not available. 26.001 DISP Set the UUT as follows: 26.001 DISP 20 dB ON 26.001 DISP RANGE 5K 26.002 DISP Adjust the UUT FREQUENCY controls for a UUT readout 26.002 DISP of 5000 Hz. 26.003 DISP Set the UUT 20 dB to OFF. 26.004 HEAD [27][91]7m[27][91]5mPlease Wait, Measurement in Progress 26.005 IEEE [@34401][D30000]MEAS:VOLT:DC?[I] 26.006 HEAD * Sine Wave Tests * #26.007 MEM* 1000 26.007 MATH mem = mem / m[19] 26.008 MATH mem = mem - m[20] 26.009 MEM/ 2 26.010 MATH mem = mem + m[20] 26.011 MEME 26.012 MEME <20 26.013 MEMC V 5.00% 5000H #! T.U.R. not calculated because System Uncertainty not available. 27.001 DISP Set the UUT as follows: 27.001 DISP 20 dB ON 27.001 DISP RANGE 50K 27.002 DISP Adjust the UUT FREQUENCY controls for a UUT readout 27.002 DISP of 50.00 KHz. 27.003 DISP Set the UUT 20 dB to OFF. 27.004 HEAD [27][91]7m[27][91]5mPlease Wait, Measurement in Progress 27.005 IEEE [@34401][D30000]MEAS:VOLT:DC?[I] 27.006 HEAD * Sine Wave Tests * #27.007 MEM* 1000 27.007 MATH mem = mem / m[19] 27.008 MATH mem = mem - m[20] 27.009 MEM/ 2 27.010 MATH mem = mem + m[20] 27.011 MEME 27.012 MEME <20 27.013 MEMC V 5.00% 50kH #! T.U.R. not calculated because System Uncertainty not available. 28.001 DISP Set the UUT as follows: 28.001 DISP 20 dB ON 28.001 DISP RANGE 500K 28.002 DISP Adjust the UUT FREQUENCY controls for a UUT readout 28.002 DISP of 500.0 KHz. 28.003 DISP Set the UUT 20 dB to OFF. 28.004 HEAD [27][91]7m[27][91]5mPlease Wait, Measurement in Progress 28.005 IEEE [@34401][D30000]MEAS:VOLT:DC?[I] 28.006 HEAD * Sine Wave Tests * #28.007 MEM* 1000 28.007 MATH mem = mem / m[19] 28.008 MATH mem = mem - m[20] 28.009 MEM/ 2 28.010 MATH mem = mem + m[20] 28.011 MEME 28.012 MEME <20 28.013 MEMC V 5.00% 500kH #! T.U.R. not calculated because System Uncertainty not available. 29.001 DISP Set the UUT as follows: 29.001 DISP 20 dB ON 29.001 DISP RANGE 5M 29.002 DISP Adjust the UUT FREQUENCY controls for a UUT readout 29.002 DISP of 5000 KHz. 29.003 DISP Set the UUT 20 dB to OFF. 29.004 HEAD [27][91]7m[27][91]5mPlease Wait, Measurement in Progress 29.005 IEEE [@34401][D30000]MEAS:VOLT:DC?[I] 29.006 HEAD * Sine Wave Tests * #29.007 MEM* 1000 29.007 MATH mem = mem / m[19] 29.008 MATH mem = mem - m[20] 29.009 MEM/ 2 29.010 MATH mem = mem + m[20] 29.011 MEME 29.012 MEME <20 29.013 MEMC V 5.00% 5000kH #! T.U.R. not calculated because System Uncertainty not available. 30.001 DISP Set the UUT as follows: 30.001 DISP 20 dB ON 30.002 DISP Set the OUTPUT LEVEL full CCW. 30.003 DISP Disconnect all equipment. 30.004 HEAD {} 30.005 HEAD {* Distortion Test *} 30.006 JMP 31.001 30.007 EVAL 31.001 DISP Connect the UUT OUTPUT via a 50 ohm termination to the 31.001 DISP HIGH INPUT on the 8903A distortion analyzer. 31.002 DISP Set the UUT as follows: 31.002 DISP RANGE 5K 31.002 DISP FREQUENCY 1000 31.002 DISP OUTPUT LEVEL midway 31.002 DISP 20 dB OFF 31.003 IEEE [@8903]M3 31.004 IEEE [@8903]LN 31.005 IEEE [@8903][I] 31.006 ACC 1pct 0.015U 31.007 MATH mem1=0 31.008 MEME 31.009 MEMC pct -1.00U 1000H #! Test Tol 1, Sys Tol 0.015, TUR 66.667 (>= 4.00). 32.001 DISP Set the UUT OUTPUT LEVEL full CCW. 32.002 DISP Disconnect all test equipment. 32.003 HEAD {} 32.004 HEAD {* Square Wave Tests *} 32.005 HEAD {* Symmetry Tests *} 32.006 HEAD * Square Wave Tests * 32.007 JMP 33.001 32.008 EVAL 33.001 DISP Set the UUT as follows: 33.001 DISP RANGE 5 33.001 DISP FREQUENCY COARSE/FINE full ccw 33.001 DISP function square 33.001 DISP OUTPUT LEVEL midpoint 33.002 DISP Connect the UUT OUTPUT via a 50 ohm termination to 33.002 DISP INPUT A on the 5335A counter. 33.003 IEEE [@5335]IN;[D2000] 33.003 IEEE [@5335]FN12; 33.004 IEEE [@5335][T0][I] 33.005 ACC 50.00percent 0.001U 33.006 MEME 33.007 MEMC percent 2.00U .5H #! Test Tol 2, Sys Tol 0.001, TUR 2000.000 (>= 4.00). 34.001 DISP Set the UUT as follows: 34.001 DISP RANGE 50 34.001 DISP FREQUENCY COARSE/FINE 50Hz 34.002 IEEE [@5335][T0][I] 34.003 ACC 50.00percent 0.001U 34.004 MEME 34.005 MEMC percent 2.00U 50H #! Test Tol 2, Sys Tol 0.001, TUR 2000.000 (>= 4.00). 35.001 DISP Set the UUT as follows: 35.001 DISP RANGE 500 35.001 DISP FREQUENCY COARSE/FINE 500Hz 35.002 IEEE [@5335][T0][I] 35.003 ACC 50.00percent 0.001U 35.004 MEME 35.005 MEMC percent 2.00U 500H #! Test Tol 2, Sys Tol 0.001, TUR 2000.000 (>= 4.00). 36.001 DISP Set the UUT as follows: 36.001 DISP RANGE 5K 36.001 DISP FREQUENCY COARSE/FINE 5000Hz 36.002 IEEE [@5335][T0][I] 36.003 ACC 50.00percent 0.001U 36.004 MEME 36.005 MEMC percent 2.00U 5000H #! Test Tol 2, Sys Tol 0.001, TUR 2000.000 (>= 4.00). 37.001 DISP Set the UUT as follows: 37.001 DISP RANGE 50K 37.001 DISP FREQUENCY COARSE/FINE 50.00KHz 37.002 IEEE [@5335][T0][I] 37.003 ACC 50.00percent 0.001U 37.004 MEME 37.005 MEMC percent 2.00U 50kH #! Test Tol 2, Sys Tol 0.001, TUR 2000.000 (>= 4.00). 38.001 DISP Set the UUT as follows: 38.001 DISP RANGE 500K 38.001 DISP FREQUENCY COARSE/FINE 100.0KHz 38.002 IEEE [@5335][T0][I] 38.003 ACC 50.00percent 0.001U 38.004 MEME 38.005 MEMC percent 2.00U 100kH #! Test Tol 2, Sys Tol 0.001, TUR 2000.000 (>= 4.00). 39.001 HEAD {} 39.002 JMP 40.001 39.003 EVAL 40.001 IEEE [TERM LF] 40.002 IEEE [@5335]FN10; 40.003 IEEE [@5335][D2000][T0][I] 40.004 MEM* 1000000000 40.005 ACC 50.00nT 0.01U 40.006 MEMC 0nT -50U risetime #! Test Tol 20, Sys Tol 0.01, TUR 2000.000 (>= 4.00). 41.001 IEEE [@5335]FN10;AS1;BS1; 41.002 IEEE [@5335][D3000][T0][I] 41.003 MEM* 1000000000 41.004 ACC 50.00nT 0.01U 41.005 MEMC 0nT -50U falltime #! Test Tol 20, Sys Tol 0.01, TUR 2000.000 (>= 4.00). 42.001 IEEE [@34401]*RST[GTL] 42.002 HEAD {} 42.003 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 8 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.